HEAT-Prober
®
PLATINUM-RTD
T
HERMOMETERS
Models
392HP,
392E, 392E-S, 392EX, 392EX-S, 392EVX,
392M, 392M-S, 392MX, 392MX-S, 392VX,
360X and 700M
INSTRUCTION MANUAL
©Copyright 2004 Wahl Instruments, Inc. Printed in USA
W-392 (Rev. M)
Vista de página 0
1 2 3 4 5 6 ... 18 19

Resumo do Conteúdo

Página 1 - PLATINUM-RTD

HEAT-Prober®PLATINUM-RTDTHERMOMETERSModels392HP,392E, 392E-S, 392EX, 392EX-S, 392EVX,392M, 392M-S, 392MX, 392MX-S, 392VX,360X and 700MINSTRUCTION MANU

Página 2

1415Models 392MX, 392MVX, 392EX, and 392EVX Models 392HP, 392M and 392EUse the Circuit Board Diagrams on these pages to locate potentiometer placement

Página 3

16176.8 Fahrenheit to Centigrade ConversionThe HEAT-Prober can be ordered as a °F or °C unit. The 392 will read out in°F unless a jumper is in place.

Página 4

1819Each HP version probe, therefore, has its specific resistor to indicate to themicroprocessor the correction to be taken.Figure 8-1. Wiring Diagram

Página 5

2021for the probe readings. Both use a common subroutine, “Get M.A.”, whichcalls “Get ND”. Control and handshaking between the converter and the com-p

Página 6

2210.0 392M TROUBLE-SHOOTING AID10.1 ScopeThis section describes basic diagnostics to find the majority of the problemswith malfunctioning 392M unit

Página 7 - 4.0 SPECIFICATIONS

2425392 SERIES TROUBLE SHOOTING PROCEDUREStep Problem Action to be Performed If Correct If Incorrect13.Set the Wahl Calibrator to 212°Fand insert into

Página 8

262712.0 CALIBRATION SERVICERecalibration service can be provided at a nominal cost, if desired. Specialthermal calibration traceable to NIST is avai

Página 9

282913.2 Calibration Procedure (See Figure 13-1)13.2.1 Equipment Required1. Digital Voltmeter, Fluke 8600 41/2 digit or equivalent.2. Decade resis

Página 10 - Models 392HP, 392M and 392E

30 31HOW TO USE TROUBLE-SHOOTING PROCEDURE: Locate the problem onthis table and proceed with the action starting with Step 1. Note, for all DVMand sco

Página 11

32234 Old Weaverville Road • Asheville, North Carolina • 28804-1228Phone 800-421-2853 • 828-658-3131 • Fax 828-658-0728 • Email: [email protected]

Página 12

iiiiiRECEIVING INSPECTIONCheck your new Wahl Portable Platinum HEAT-Prober®Thermometer uponreceiving to assure that:1. It has not been physically dama

Página 13 - Monitor/Peak

iv 1TABLE OF CONTENTS (Continued)8.3 Intrinsically Safe for Hazardous Location(-S Versions) 178.4 Rate of Change with Hot and ColdPeak Hold (Option R)

Página 14

232.0 OPERATION (See Figure 2-1)Insert probe into jack at top. Turn on the switch at the left. The switch is 3-position; the lower detent is marked “

Página 15 - METER OPTIONS

452.4 The BatteriesThe 392HP, 392E, 392EX, 392M, and 392MX HEAT-Prober are powered withfive size AA rechargeable nickel-cadmium batteries, 1.2 volts

Página 16

673.0 PROBESProbes compatible with the 392 HEAT-Prober are constructed utilizing a wirewound Platinum RTD sensor with an R0 of 100 Ohms wired in a 3-

Página 17

89392 SERIES METER SPECIFICATIONSModel392HP392E 392EX392E-S392EX-S392EVXRange-290 to 1450°F-180 to 788°C-290 to 1450°F-180 to 788°C-290 to 1450°F-180

Página 18 - ROUBLE SHOOTING PROCEDURE

10115.0 MAINTENANCE/REPAIRThe HEAT-Prober instrument may be recalibrated, or adjusted for a particularprobe in the field by following the directions

Página 19 - W-392 (Rev. M) 11/04

1213Keep decreasing the supply voltage until 5.5 volts is reached, as measuredwith a digital DVM. The meter should maintain correct operation even tho