TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
VWR Scientific 34100–674 pH Meter
MFG. MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20SC–42
Page 2 of 2
4.3 PH Calibration Tests
4.3.4 TI control(s):
FUNCTION sw. to PH
Adjust STANDARDIZE
for TI display of 7.00 adjust
4.3.6 TI control(s):
Adjust SLOPE for TI
display of 4.01 ck ( ) NA adjust
4.3.8 pH measurement pH) (pH)
4.01 3.98 to 4.04
4.3.10 pH measurement (pH) ck ( ) NA (pH)
10.01 9.98 to 10.04
4.4 Temperature Indication Tests
4.4.1 TI control(s):
FUNCTION sw. to TEMP
4.4.4 TI temperature indication 25C +
5 C of std. ind.
NAVAIR 17–20SC–42
12