10/08
172 Years of Continued Innovation
ISO9001:2000 CERTIFIED
Wahl
®
DST600
Solution
Calibration Services Available
Complete Calibration Services Available
FDA Proposed Rule 21CFR Part 113 Wahl DST600 Solution
Items 1 and 2 are from the Preamble.
4b Ambient Temperature variations Ambient Temperature coefficient of no
greater than .003°/°. This equates to no
more than .03° change in reading for a 10°
ambient temperature change.
4c Vibration Tested in accordance with MIL STD 202G
Method 201A extended to 60Hz vibration: 10
to 60 Hz.
4d Each temperature-indicating device and refer-
ence device shall have a tag, seal, or other
means of identity.
DST600's are clearly marked with unique
serial numbers visible from the front panel.
Serial numbers are also electronically stored
in internal memory and cannot be tampered
with by the user. Custom user identification
numbers may be added by user.
5 113.40 (ii) Written documentation of the accuracy of
the temperature-indicating device and the
reference device shall be established and
maintained.
Processor is responsible for establishment
and maintenance of documents. All NIST
traceable certifications performed by Wahl
are provided in writing and include test data
showing "as received" analysis and "as
returned" accuracy of the instrument. Copies
of these certifications are maintained in our
facility for the life of the instrument if you
have Palmer Wahl certify your DST600 every
year.
6 113.40 (A) Documentation of the accuracy of the tem-
perature-indicating device shall include a ref-
erence to the tag, seal, or other means of
identity used by the processor to identify the
temperatur
e-indicating device, the name of
the manufacturer of the temperature-indicat-
ing device, the identity of the reference
device used for the accuracy test and of
equipment and procedures used to adjust or
calibrate the temperature-indicating device,
the date and results of each accuracy test,
the name of the person or facility that per-
formed the accuracy test and adjusted or cal-
ibrated the temperature-indicating device,
and the date of the next scheduled accuracy
test.
Wahl’s NIST traceable certificate includes all
of the requirements of this paragraph.
W
ahl
recommends you return product to Palmer
Wahl for re-calibration.
If pr
ocessors per
form their own testing,
compliance to this paragraph is required.
Wahl's calibration software is simple, fast
and easy to use.